1BThe collimator provides directional information by accepting photons travelling along specific trajectories.
2AParallel-hole collimator resolution improves as the distance between the patient and collimator decreases.
3BTotal counts are the product of count rate and acquisition time. If the count rate decreases by half, the acquisition time must be doubled to obtain approximately the same total counts and preserve image count statistics.
4CThe photopeak energy window preferentially accepts primary photons and rejects many photons that have undergone Compton scattering.
5ADetector non-uniformities are reproduced at different projection angles and may form ring artefacts after SPECT reconstruction.